Caloric value test of Animal feeding stuffs & animal products2026-03-23  view:27

    Standards: BS EN ISO9831 and GB/T 45104


    Sample Requirements

1 Particle size ≤ 1.0 mm

2 Sample should be Packed with Lens Paper


Test Procedure

Add a specified amount of distilled water (BS EN ISO9831:5ml , GB/T 45104:10ml) to oxygen bomb ,Weigh approximately 0.9-1.1 g of sample into lens paper of known mass and calorific value, wrap it, and place it in a metal crucible. Install the ignition wire or ignition cotton thread, and close the oxygen bomb lid. Place the oxygen bomb into the calorimeter. Input the sample weight and lens paper weight, and the instrument will automatically complete the calorific value determination and result calculation.

    Test Conditions

· Oxygen pressure: 3.0 MPa

· Oxygenation time: ≥15s

Ambient temperature : 23℃                 

Jacket temperature : 28℃

Bucket temperature : 26℃               

Water tank temperature : 24℃ 


    Instrument : SDAC1200

Test Results

Sample Name

Sample mass (g)

Qb,ad

 ( MJ / kg )

Mean

 ( M J/ kg )

Deviation ( M J/ kg )

Precision requirements 

( M J/ kg )

Feed 1

0.5117

16.9

16.8

0.1

0.84

Feed 1

0.5170

16.8

0.0

Feed 2

0.5054

17.6

17.6

0.0

0.88

Feed 2

0.5108

17.7

0.1

 